NMR Experiment Guide

UC Chemistry Nuclear Magnetic Resonance Facility: Table of NMR experiments currently available in automation mode.

NMR Experiment Guide
Parameter Set Default Time Description Special Instructions Pulse Sequence Instrument(s) 
PROTON 1min30s 1D 1H, uses 30° pulse for reliable integrals. Change NS to increase signal if desired. Switch to night queue if increasing length over 20 min. zg30 NEO400, AV400, AV500
C13CPD32 1min56s 1D 13C, 32 scans. Uses decoupling during delay and acquisition for increased sensitivity. Provides OK S/N for ~25 mg samples. Carbons without attached protons will be weaker. Change NS to increase signal if desired. Do not set length longer than 30 min on NEO400. zgpg30 NEO400, AV400, AV500
C13CPD 59min Same as C13CPD32 but runs for 1 hr. Night queue only. Provides OK S/N for ~6 mg samples.  Check available night queue time before submitting. zgpg30 NEO400, AV400, AV500
C13DEPT135 17min55s 1D 13C DEPT135. Sensitivity enhanced and multiplicity edited. Carbons without attached protons are not observed. CH and CH3 carbons should be positive and CH2 carbons should be negative (or backwards if phased upside-down). Night queue only. Check available night queue time before submitting. deptsp135 NEO400, AV400, AV500
WATERSUP1 2min17s 1D 1H with largest peak  automatically suppressed (usually should be solvent). Will not suppress multiple signals (such as for ETOH or THF). 90° pulse width. Do not use unless there is a large, undesirable peak in spectrum. Peaks near the large peak may also be affected. noesygppr1d NEO400, AV400, AV500
PROB11DEC 54s 1D 1H with 11B decoupled.  Set O2p right on resonance with 11B signal. zgig NEO400, AV400, AV500
PROF19DEC 54s 1D 1H with 19F decoupled.  Set O2p right on resonance with 19F signal. zgig NEO400, AV400
PROP31DEC 54s 1D 1H with 31P decoupled.  Set O2p right on resonance with 31P signal. zgig NEO400, AV400, AV500
SAPPHIRE_pureshift1H 6min41s 1D 1H (reconstructed) with 1H-1H couplings removed (pure shift). Geminal protons will still be doublets. Provides OK S/N for ~6 mg samples.  Note the much-lower sensitivity compared to PROTON. HSQC_PURESHIFT may provide better sensitivity. sapphire_ZS_AG NEO400
PROTON_BIGSW 1min17s Same as PROTON but the window is 50 ppm wide and centered at -5 ppm.   zg30 NEO400, AV400, AV500
PROTONT1 25min14s Inversion-recovery measurement to determine 1H T1s.  Data can be analyzed with the command "t1guide" in Topspin.  t1ir NEO400, AV400, AV500
F19 35s 1D F19, no decoupling. Probe background around -170 ppm can be removed with linear prediction if desired. zg NEO400, AV400
F19CPD 35s 1D F19, 1H decoupling during acquisition. Probe background around -170 ppm can be removed with linear prediction if desired. zgig NEO400, AV400
F19ECHO 4min38s 1D F19 with background-suppressing echo, no decoupling, and a 30° pulse width for more accurate integrals. Collect a F19 spectrum as well. F19 signals with strong couplings (CF2H, geminal CF2s) can be badly distorted with the echo. Do not rely on this spectrum alone for quantification. zgsezr NEO400, AV400
F19ECHO_IG 4min38s 1D F19 with background-suppressing echo, 1H decoupling during acquisition, and a 30° pulse width for more accurate integrals. Collect a F19CPD spectrum as well. F19 signals with strong couplings (CF2H, geminal CF2s) can be badly distorted with the echo. Do not rely on this spectrum alone for quantification. zgigsezr NEO400, AV400
F19T1IR 9min4s Inversion-recovery measurement to determine 19F T1s.  Data can be analyzed with the command "t1guide" in Topspin.  t1irig NEO400, AV400
P31 1min31s 1D 31P, uses 30° pulse for reliable integrals. No decoupling. Change NS to increase signal if desired. Switch to night queue if increasing length over 20 min. zg30 NEO400, AV400, AV500
P31CPD 51s 1D 31P, uses 30° pulse for reliable integrals. Uses decoupling during delay and acquisition for increased sensitivity.    zgpg30 NEO400, AV400, AV500
P31_CRP20 1min32s 1D 31P, uses 30° pulse for reliable integrals. Uses decoupling during delay and acquisition for increased sensitivity. Uses chirped decoupling to decouple larger bandwidth of protons (from about 10 ppm to -30 ppm). Use this if you need to decouple protons both in the normal 0-8 ppm range but also down to about -30 ppm.   zgpg30 NEO400, AV400, AV500
COSYPH 1hr27min 2D 1H-1H COSY (COrrelation SpectroscopY). Night queue only.    cosyph NEO400, AV400, AV500
COSYPHPR 1hr28min 2D COSYPH with automatic solvent suppression. Does not suppress multiple signals. Night queue only. Set O1P to the position of your solvent peak (only on AV400 and AV500). Only use this if there is a large, undesirable peak in spectrum. Peaks near the large peak may also be affected. cosyphpr NEO400, AV400, AV500
NOESYGPPHSW 49min53s 2D NOESY (Nuclear Overhauser Effect SpectroscopY). Measures 1H-1H through-space contacts and chemical exchange. Default mixing time D8 of 0.3 sec (but can be modified). With positive diagonal peaks, negative cross peaks indicate NOE or chemical exchange. Night queue only.   noesygpph NEO400, AV400, AV500
ROESYPHSW_N 47min46s 2D ROESY (ROtating frame noESY). Measures 1H-1H through-space contacts and chemical exchange like NOESY. With positive diagonal peaks, positive cross peaks indicate NOE contacts, and negative cross peaks indicate chemical exchange. Night queue only. Can be better than NOESY for molecules of approximately 1000 Da. roesyphpp.2 NEO400, AV400, AV500
HSQCEDETGPSISP.2_ADIA 55min51s 2D 1H-13C HSQC (Heteronuclear Single Quantum Coherence). Multiplicity-edited so CH and CH3 peaks are positive and CH2 are negative.   hsqcedetgpsisp.2 NEO400, AV500
HSQC_PURESHIFT 1hr7min 2D 1H-13C HSQC (Heteronuclear Single Quantum Coherence). Multiplicity-edited so CH and CH3 peaks are positive and CH2 are negative. 1H signals are pure shift (1H-1H couplings removed) using BIRD decoupling for improved signal and resolution. Night queue only.   hsqcedetgpsisp2.3_bbhd NEO400, AV500
HMBCETGPL3ND_13C 1hr25min 2D 1H-13C HMBC (Heteronuclear Multiple Bond Correlation). Gives 2+ bond correlations between 13C and 1H. 3-bond correlations are often stronger than 2- and 4-bond. Night queue only.   hmbcetgpl3nd NEO400, AV400, AV500
1H_19F_HMBC_2D 1hr25min 2D 1H-19F HMBC (Heteronuclear Multiple Bond Correlation). Night queue only. You should set the 1sw (window) and o2p (center frequency) to cover the range of signals in your 19F 1D but not much larger. hmbcetgpl3nd NEO400, AV400
1H_19F_HOESY_2D 3hr7min 2D 1H-19F HOESY (Heteronuclear Overhauser Effect SpectroscopY). Gives 19F-1H through-space contacts. Night queue only. You should set the 1sw (window) and o2p (center frequency) to cover the range of signals in your 19F 1D but not much larger. hoesyetgpsp.2.nodec NEO400
F19COSY 1hr29min 2D 19F-19F COSY (COrrelation SpectroscopY). Night queue only.  Set 1sw, 2sw (window), o1p and o2p (center frequencies) to cover range of 19F and not larger. Be careful--19F-19F J couplings are sometimes very different from expected based on 1H-1H couplings due to geometry differences.  cosygpqf NEO400
F19NOESY 4hr9min 2D NOESY (Nuclear Overhauser Effect SpectroscopY). Measures 19F-19F through-space contacts. Night queue only. Set 1sw, 2sw (window), o1p and o2p (center frequencies) to cover range of 19F and not larger. noesyetgp.sp NEO400
1H_31P_HMBC_2D 42min50s 2D 1H-31P HMBC (Heteronuclear Multiple Bond Correlation). Night queue only. You should set the 1sw (window) and o2p (center frequency) to cover the range of signals in your 31P 1D but not much larger. hmbcetgpl3nd NEO400, AV400, AV500
DOSY 20min36s Diffusion Ordered SpectroscopY. Separates 1H NMR signals by diffusion coefficients. Good for mixtures and for determining rough molecular weights.  Use a dilute solution and a medium viscosity solvent like H2O. Analysis is best done using Topspin, not MNova. ledbpgp2s NEO400, AV500
DOSY_CC 20min42s Like DOSY but with convection compensation for more accurate results, especially for low viscosity solvents. Use a dilute solution and a medium viscosity solvent like H2O. Analysis is best done using Topspin, not MNova. dstebpgp3s NEO400, AV500
PROJECTED_DOSY_CC 17min23s Like DOSY_CC but with compensation for chemical exchange effects. Use a dilute solution and a medium viscosity solvent like H2O. Analysis is best done using Topspin, not MNova. projecteddosy_cc NEO400
2Hzg 7min29s 1D 2H, no decoupling. Prepare sample in a protonated solvent! zg2h NEO400, AV500
7Li_ZG 12min6s 1D 7Li, no decoupling.   zg NEO400, AV500
B11ZG 5min3s 1D 11B, no decoupling.   zg NEO400, AV400, AV500
B11_BS 1min5s 1D 11B, no decoupling. Background suppression.   zgbs NEO400, AV400, AV500
B11_IGBS 1min6s 1D 11B, 1H decoupling during acquisition. Background suppression.   zgigbs NEO400, AV400, AV500
N14 6min8s 1D 14N, no decoupling.   zg NEO400
N15 6min30s 1D 15N, no decoupling.   zg NEO400, AV400, AV500
N15IG 3min37s 1D 15N, 1H decoupling during acquisition.   zgig NEO400, AV400, AV500
Mg25_zg 1min5s 1D 25Mg, no decoupling.   zg NEO400
Mg25_BS 1min5s 1D 25Mg, no decoupling. Background suppression.   zgbs NEO400
AL27ND 1min23s 1D 27Al, no decoupling.    zg NEO400, AV400, AV500
SI29IG 13min22s 1D 29Si, 1H decoupling during acquisition.   zgig NEO400, AV500
SE77ZG 1min43s 1D 77Se, 1H decoupling during acquisition and a 30° pulse width for more accurate integrals.   zgig30 NEO400, AV400, AV500
SN119IG   48s 1D 119Sn, 1H decoupling during acquisition and a 30° pulse width for more accurate integrals.   zgig30 NEO400, AV400, AV500